LS-1-CAL Radiometric Calibration
Standards
The LS-1-CAL and HL-2000-CAL Calibrated Tungsten Halogen Light Sources calibrate the absolute spectral response of a radiometric system. With one of these NIST-traceable light sources and our software, you can determine known absolute intensity values at wavelengths from 300-1050 nm.
NIST-traceable Calibration from 300-1050 nm. Provides absolute spectral intensity in µW/cm2/nm at the fiber port
Sources can be calibrated specifically for a bare fiber, a fiber plus cosine corrector, or an integrating sphere
Calibration Certificate. Calibration data is provided in paper and electronic formats.
The LS-1-CAL and HL-2000-CAL are specifically designed for calibrating the absolute spectral response of a complete system consisting of a spectrometer and an optical fiber and/or a CC-3-UV Cosine Corrector. The calibration data for both the LS-1-CAL and the HL-2000-CAL includes absolute intensities for wavelengths between 300–1050 nm.
Calibrated for Use with Integrating SphereThe LS-1-CAL-INT is designed for calibrating the absolute spectral response of your system when using the FOIS-1 Fiber Optic Integrating Sphere as your sampling optic. The LS-1-CAL-INT comes with a PTFE diffuser plug that fits snugly into the sample port of the FOIS-1 Integrating Sphere to measure absolute spectral intensities of LEDs and other emission sources. The HL-2000-CAL-ISP is designed for calibrating the absolute spectral response of your system when using the ISP-50-8-I Integrating Sphere as your sampling optic.
Included with the LampsEach one of these radiometric sources comes with a regulated 12 VDC power supply. Also included is a calibration certificate and a CD with data files for use with our software.
Other Calibration ServicesThe LS-1-CAL provides 50 hours of operation before recalibration is necessary. We recalibrate these lamps in-house. (For more information on LS-1-RECAL, contact an Applications Scientist.) For those with spectroradiometric systems working below the 300-1050 nm range of the LS-1-CAL and HL-2000-CAL, see our DH-2000-CAL.
PricingItem Code | Description | Price |
LS-1-CAL | Radiometric Calibration Standard, 300-1050 nm | $866 |
LS-1-CAL-INT | Radiometric Calibration Standard, 300-1050 nm | $866 |
LS-1-CAL-EXT | Upgrade for extended range. | $250 |
LS-1-RECAL | Radiometric Re-calibration Service for the LS-1-CAL and LS-1-CAL-INT | $199 |
HL-2000-CAL | Radiometric Calibration Standard, 300-1050 nm | $993 |
HL-2000-CAL-ISP | Radiometric Calibration Standard, 300-1050 nm | $1150 |
HL-2000-RECAL | Radiometric Re-calibration Service for the HL-2000-CAL and HL-2000-CAL-ISP | $199 |
LS-1-CAL (Serial Number LSC712) Calibration Using CC-3
Spectral range (calibrated): | 300-1050 nm (calibrated) |
Power consumption: | 600 mA @ 12 VDC |
Power output: | 6.5 watts |
Bulb life: | 900 hours (recommend recalibration after 50 hours of use) |
Recalibration: | Required after ~50 hours of operation |
Bulb color temperature: | 3100 K |
Output to bulb: | 5 volts/1.3 amps |
Output regulation: | 0.2% voltage |
Time to stabilized output: | ~30 minutes |
Connector: | SMA 905 for fiber; 6.35-mm barrel for cosine corrector; PTFE plug for integrating sphere |